Hach Sension+ EC7 Benchtop Conductivity Meter Kit with Electrode Stand, Magnetic Stirrer, Accessories and 5070 Electrode Cell

Code: WAT4584 24

The Hach SensION+ EC7 benchtop conductivity meter is designed to be an 'all-in-one' working station, and comes complete with a basic laboratory conductivity meter, an integrated electrode holder, a...

read more

ADD
£889.30 EACH List Price

The Hach SensION+ EC7 benchtop conductivity meter is designed to be an 'all-in-one' working station, and comes complete with a basic laboratory conductivity meter, an integrated electrode holder, and magnetic stirrer. Guided menu navigation makes use fast and simple, and it is ready to use for a wide variety of applications. Also includes standard solutions (147 & 1413 µS/cm and 12.88 mS/cm), calibration flasks with magnetic stirring bar and user manual. WAT4584 includes electrode 5070 for general applications.

  • Comprehensive lab station
  • Incorporated electronics adapt to platinum and titanium cells
  • Calibration with 1,2 or 3 standards
Accuracy
±1 digit
AC power adapter & USB/DC adapter
Yes
Calibration intervals/reminders
Reminder 1 hour to 99 days
Conductivity accuracy
≤ 0.5 % of measurement range
Conductivity measurement
Linear Coefficient: 0.00 - 9.99 %/°C
Reference Temperature: 20 °C or 25 °C
Conductivity measurement range
0.01 µS/cm - 500 mS/cm
Conductivity resolution
0.01 µS/cm - 1 mS/cm (depending on the measurement range)
Data logger
No
Data storage
No
Dimensions (H x W x D)
110 mm x 350 mm x 201 mm
Display
128 x 64 pixel
Display type
Backlit Graphical LCD
Electrode stand
Yes
Environmental conditions
< 80 % relative humidity (non-condensing)
0 - 40 °C (41 - 104 °F)
Inputs
Conductivity probe with built-in Pt1000 sensor
Magnetic stirrer
Interface languages
English, German, French, Spanish, Italian, Portuguese
Parameter
Conductivity, Salinity, Resistivity, Temperature
Power
External power supply 100-240 V, 0.4 A, 47-63 Hz with plug adapters
Resistivity measurement
0 Ω - 500 MΩ
Resolution
Depending on the measurement range
EC: 0.01 µS/cm - 1 mS/cm depending on range
Salinity/Resistivity: dependent on range
≤ 0.5 %
Salinity measurement range
5.85 mg/L - 311.1 g/L NaCl
Salinity resolution
0.01 mg/L - 0.1 g/L NaCl, depending on the measurement range
Temperature accuracy
≤ 0.2 °C
Temperature compensation
Manual or Automatic
Temperature measurement range
-20.0 - 150.0 °C (-4.0 - 302.0 °F)
Temperature resolution
0.1 °C (0.1 °F)
Weight
1100 g